Device product
Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.
Z-1674-2020
Product summary
- Event
- Event 85137
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3,472 units
- Official record key
device-enforcement:Z-1674-2020
Official wording
Reason: Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.
Code information: Lot number 20201103T, Expiration Date: 11/3/2020
Distribution pattern: Worldwide distribution - US Nationwide distribution in the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MI, MS, NE, NY, NC, ND, OH, OK, PA, SD, TN, TX, UT, WV, WI, WY and countries of Canada, China, Japan, Paraguay.
Derived failure modes
-
Unknown
Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.