Device product
Philips Sterilizable Defibrillator Internal Paddles - Internal paddles are an accessory to the compatible monitor/defibrillator Switchless: M1741A, M1742A, M1743A, M1744A Switched: M4741A, M4742A, M4743A, M4744A - Product Usage: An electrical conductor used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.
Z-2134-2020
Product summary
- Event
- Event 85571
- Status
- Terminated
- Classification
- Class II
- Quantity
- 21,683 Internal Paddle Sets.
- Official record key
device-enforcement:Z-2134-2020
Official wording
Reason: Internal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instructions for Use (IFU) to perform routine operational checks are followed. The instructions for use rely on these checks rather than specifying a maximum number of sterilization cycles to determine when the paddles should be discarded. A damaged Internal Paddle may not be able to deliver therapy.
Code information: All lots
Distribution pattern: Worldwide distribution - US Nationwide distrubtion and the countries of Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Congo (Democratic Republic of the), Costa Rica, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Haiti, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Lao People's , Democratic Republic Latvia, Lebanon, Libya, Macao, Macedonia, Malaysia, Maldives, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman Pakistan, Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania Russian Federation, Saudi Arabia, Serbia, Singapore Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
Derived failure modes
-
Unknown
Internal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instructions for Use (IFU) to perform routine operational checks are followed. The instructions for use rely on these checks rather than specifying a maximum number of sterilization cycles to determine when the paddles should be discarded. A damaged Internal Paddle may not be able to deliver therapy.