Device product
MRIdian Linac Radiation Therapy System Model 10000 and 20000
Z-0676-2020
Product summary
- Firm
- Viewray, Inc.
- Event
- Event 84303
- Status
- Terminated
- Classification
- Class II
- Quantity
- 24 Systems
- Official record key
device-enforcement:Z-0676-2020
Official wording
Reason: Issue with transferring treatment regiment from one system to another. Adequate instructions for transferring treatment plan(s) are not provided within the operator's manual or within the software operating system.
Code information: Serial Numbers: 100, 101, 102, 103, 104, 105, 107, 108, 112, 113, 201, 202, 203, 204, 205, 206, 209, 211, 213, 214, 217, 221, 223, 225 Device Listing Number: D158423 and D297999
Distribution pattern: US - CA, FL, IL, MI, MO, NY, and WI OUS: Denmark, France, Germany, Italy, Japan, Netherlands, Turkey, South Korea, and UAE
Derived failure modes
-
Unknown
Issue with transferring treatment regiment from one system to another. Adequate instructions for transferring treatment plan(s) are not provided within the operator's manual or within the software operating system.