Device product
iChem¿VELOCITY Urine Chemistry System-
Z-1944-2020
Product summary
- Event
- Event 85423
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1557 units
- Official record key
device-enforcement:Z-1944-2020
Official wording
Reason: A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
Code information: Catalog #700-7176-001 and #700-7177-001, All part numbers
Distribution pattern: US Distribution: Nationwide Foreign Distribution: Worldwide
Derived failure modes
-
Unknown
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.