Skip to content
Recall Observatory FDA recall evidence

Device product

Remisol Advance running on hardware with Windows XP, Windows 7, Windows Server 2003 and Windows Server 2008 operating systems

Z-1430-2020

August 15, 2019

Class II

Product summary

Firm
Normand-Info S.A.S.U.
Event
Event 83687
Status
Terminated
Classification
Class II
Quantity
6479
Official record key
device-enforcement:Z-1430-2020

Official wording

Reason: A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.

Code information: All serial numbers

Distribution pattern: U.S.: NY, AZ, FL, NJ, MA, OH, CA, IL, NC, OK, IN, KY, OR, MI, WI, PA, UT, GA, VA, TX, NE, SD, IA, MN, MD, DE, SC, MT, WV, LA, MO, TN, NM, NV, HI, WA, WY, AL, AR, MS, RI, ID, ME, DC, ND, AK, CO, CT, KS, NH, VT

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.