Skip to content
Recall Observatory FDA recall evidence

Device product

ADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA¿ Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).

Z-1539-2020

February 13, 2020

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 85037
Status
Terminated
Classification
Class II
Quantity
5666 US and 52202 OUS
Official record key
device-enforcement:Z-1539-2020

Official wording

Reason: Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy

Code information: ADVIA CH TBIL_2 (40 mL) 492094 (01)00630414513072(10)492094(17)20210828 ADVIA CH TBIL_2 (40 mL) 486525 (01)00630414513072(10)486525(17)20210728 ADVIA CH TBIL_2 (40 mL) 478149 (01)00630414513072(10)478149(17)20210428 ADVIA CH TBIL_2 (40 mL) 472677 (01)00630414513072(10)472677(17)20210328 ADVIA CH TBIL_2 (40 mL) 469083 (01)00630414513072(10)469083(17)20210128 ADVIA CH TBIL_2 (40 mL) 462464 (01)00630414513072(10)462464(17)20201128 ADVIA CH TBIL_2 (40 mL) 455486 (01)00630414513072(10)455486(17)20200828 ADVIA CH TBIL_2 (40 mL) 452851 (01)00630414513072(10)452851(17)20200728 ADVIA CH TBIL_2 (40 mL) 447317 (01)00630414513072(10)447317(17)20200628 ADVIA CH TBIL_2 (40 mL) 436498 (01)00630414513072(10)436498(17)20200328 ADVIA CH TBIL_2 (70 mL) 492102 (01)00630414513058(10)492102(17)20210828 ADVIA CH TBIL_2 (70 mL) 492111 (01)00630414513058(10)492111(17)20210828 ADVIA CH TBIL_2 (70 mL) 486533 (01)00630414513058(10)486533(17)20210728 ADVIA CH TBIL_2 (70 mL) 478154 (01)00630414513058(10)478154(17)20210428 ADVIA CH TBIL_2 (70 mL) 478159 (01)00630414513058(10)478159(17)20210428 ADVIA CH TBIL_2 (70 mL) 493710 (01)00630414513058(10)493710(17)20210428 ADVIA CH TBIL_2 (70 mL) 490263 (01)00630414513058(10)490263(17)20210328 ADVIA CH TBIL_2 (70 mL) 490264 (01)00630414513058(10)490264(17)20210328 ADVIA CH TBIL_2 (70 mL) 469088 (01)00630414513058(10)469088(17)20210128 ADVIA CH TBIL_2 (70 mL) 469094 (01)00630414513058(10)469094(17)20210128 ADVIA CH TBIL_2 (70 mL) 462470 (01)00630414513058(10)462470(17)20201128 ADVIA CH TBIL_2 (70 mL) 462477 (01)00630414513058(10)462477(17)20201128 ADVIA CH TBIL_2 (70 mL) 455496 (01)00630414513058(10)455496(17)20200828 ADVIA CH TBIL_2 (70 mL) 455501 (01)00630414513058(10)455501(17)20200828 ADVIA CH TBIL_2 (70 mL) 452858 (01)00630414513058(10)452858(17)20200728 ADVIA CH TBIL_2 (70 mL) 452863 (01)00630414513058(10)452863(17)20200728 ADVIA CH TBIL_2 (70 mL) 452870 (01)00630414513058(10)452870(17)20200728 ADVIA CH TBIL_2 (70 mL) 447325 (01)00630414513058(10)447325(17)20200628 ADVIA CH TBIL_2 (70 mL) 447332 (01)00630414513058(10)447332(17)20200628 ADVIA CH TBIL_2 (70 mL) 436510 (01)00630414513058(10)436510(17)20200328 ADVIA CH TBIL_2 (70 mL) 436515 (01)00630414513058(10)436515(17)20200328

Distribution pattern: Nationwide Foreign: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom, Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Azerbaijan, Bahamas, Bahrain, Bangladesh, Benin, Bosnia Herzegovina, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Ecuador, Egypt, French Guinea, Georgia, Guadeloupe, Hong Kong, India, Indonesia, Iraq, Iran, Israel, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Macedonia, Malawi, Malaysia, Mali, Mexico, Montenegro, Morocco, Mozambique, Myanmar, New Zealand, Nigeria, Oman, Pakistan, Peru, Philippines, Qatar, Republic if Yemen, Republic Korea, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, South Africa, Swaziland, Taiwan, Tanzania, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Ukraine, Uruguay, Venezuela, Vietnam, Zimbabwe

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy