C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Z-0613-2020

October 21, 2019

Class II

Product summary

Firm
Integra LifeSciences Corp.
Event
Event 84216
Status
Terminated
Classification
Class II
Quantity
820
Official record key
device-enforcement:Z-0613-2020

Official wording

Reason: The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Code information: US Ops Manual P/N 60905769

Distribution pattern: US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.