Expression MR400 MRI Patient Monitoring System Model Number: 866185 Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00 - Product Usage: The Expression Model MR400 MRI Patient Monitoring Systems is intended for use by healthcare professionals to monitor vital signs of patients undergoing MRI procedures and to provide signals for the synchronization of the MRI scanner.

Z-1616-2020

February 20, 2020

Class II

Product summary

Firm
Philips North America, LLC
Event
Event 85090
Status
Terminated
Classification
Class II
Quantity
2214 units
Official record key
device-enforcement:Z-1616-2020

Official wording

Reason: Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00

Code information: Software version number: 01.03.00, 01.04.00, 01.05.00, and 01.07.00

Distribution pattern: Worldwide distribution - US Nationwide distribution and countries of Utd Arab Emir, Albania, Austria, Australia, Azerbaijan, Bulgaria, Brazil, White Russia, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Israel, India, Iceland, Italy, Jamaica, Japan, Kyrgyzstan, Malaysia, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Uzbekistan, Viet Nam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Menu selections for users to access the oxygen (O2) sensor calibration were changed and not incorporated into the device s Instructions for Use. As a result, the instructions for performing the O2 sensor calibration are incorrect for customers performing this task with an MR400 device with software revisions: 01 .03.00, 01.04.00, 01 .05.00 and 01 .07.00