Skip to content
Recall Observatory FDA recall evidence

Device product

Pharos Excimer Laser, Model: EX-308 - Product Usage: The Pharos EX-308 Excimer Laser System is designed to treat the skin disorders psoriasis, vitiligo, atopic dermatitis, and leukoderma using phototherapy. The excimer laser system produces collimated 308 nm radiation, within the invisible region of the electromagnetic spectrum. The UV radiation falls within the range of ultraviolet B (UVB), which is used to treat photo-responsive skin disorders.

Z-1631-2020

August 23, 2019

Class II

Product summary

Firm
Ra Medical Systems, Inc.
Event
Event 85200
Status
Terminated
Classification
Class II
Quantity
10
Official record key
device-enforcement:Z-1631-2020

Official wording

Reason: At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause overtreatment, which could result in strong erythema, and possibly cause blisters.

Code information: All serial numbers.

Distribution pattern: U.S. Nationwide distribution in the states of CA, FL, OR, MS, NY.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software error