Skip to content
Recall Observatory FDA recall evidence

Device product

Neo-Vent Infant T-Piece Resuscitation Circuit;Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0415, Qty:10, Length: 60" (1.5cm), Rx Only.

Z-2434-2020

January 15, 2019

Class II

Product summary

Firm
Westmed, Inc.
Event
Event 85576
Status
Terminated
Classification
Class II
Quantity
625 cases
Official record key
device-enforcement:Z-2434-2020

Official wording

Reason: A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Code information: 101118T42, 101118T44, 102918T01, 080618T02, 080618T03 , 101118T43, 102918T02.

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.