Device product

Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031,

Z-1452-2020

August 22, 2019

Class II

Product summary

Firm
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Event
Event 84881
Status
Terminated
Classification
Class II
Quantity
669 trollys
Official record key
device-enforcement:Z-1452-2020

Official wording

Reason: The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.

Code information: Concert & Basic Shower trolleys Serial Numbers: P0512263 to P0551621, P0279164, SEE0643339, SEE0605124,P0249502, SEE045073, SEE0520032, 1106000338, SEE0348065, P0422951, P0232526

Distribution pattern: US: CA, CO, CT, GA, MA, MD, MI, NC, NJ, NY, OH, OK, TX, VA, WI OUS: Australia Austria, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iceland, India, Ireland, Italy, Japan, Korea, Luxembourg, Netherlands, Netherlands/CW, New Zealand, Norway, Philippines, Poland, Romaina, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, United Arab Emirates, United Kingdom

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.