Skip to content
Recall Observatory FDA recall evidence

Device product

Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. PC3688 CE 0482, Adhesif Film Ref. 38.20.15 CE 0459, EN Navigation Kit. Compatible with linear instruments of diameter ranging from 0.5 mm to 3.0 mm (25G to 11G), Sterile, Rx Only,

Z-0838-2020

August 01, 2019

Class II

Product summary

Firm
Imactis
Event
Event 84460
Status
Ongoing
Classification
Class II
Quantity
122 kits
Official record key
device-enforcement:Z-0838-2020

Official wording

Reason: The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.

Code information: Lot # 18120001/Exp. 2020-04-01 and Lot # 19040001/Exp. 2021-01-22

Distribution pattern: US: WI & IL

Derived failure modes

  • Microbial contamination

    reason.microbial_contamination · v1.0.0

    microbial contamination