Device product
ARCHITECT HAVAB G Reagent Kit. List Number 6L27 25.
Z-1728-2020
Product summary
- Event
- Event 85238
- Status
- Terminated
- Classification
- Class II
- Quantity
- 7,347 kits
- Official record key
device-enforcement:Z-1728-2020
Official wording
Reason: An erroneous concentration value for hepatitis A virus was utilized during manufacture, leading to a decreased amount of antigen used for the impacted lots.
Code information: List Number 6L27 25. Lot No: 03429BE00, 06172BE00, 08073BE00, 10353BE00. UDI: (01)00380740011185 (17)200320(10)03429BE00, (01)00380740011185 (17)200604(10)06172BE00, (01)00380740011185 (17)200819(10)08073BE00, (01)00380740011185 (17)201023(10)10353BE00.
Distribution pattern: US nationwide including the states of AL, AR, AZ, CA, CO, CT DE, FL, GA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM NV, NY, OH, OK, OR, PA, RH, SC, TN, TZ, UT, VA, WA, WI, WV AND PR.
Derived failure modes
-
Unknown
An erroneous concentration value for hepatitis A virus was utilized during manufacture, leading to a decreased amount of antigen used for the impacted lots.