Device product
Panther Fusion Extraction Reagent-S, Ref PRD-04331, IVD, CE, Rx Only consisting of FCR-S Fusion Capture Reagent - S, Nucleic Acid in a buffered solution and FER-S Fusion Enhancer Reagent-S (Lithium Hydroxide Solution), UDI: 15420045509214
Z-2163-2020
Product summary
- Firm
- Hologic, Inc
- Event
- Event 85346
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 47 kits
- Official record key
device-enforcement:Z-2163-2020
Official wording
Reason: Reagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to invalid runs. This could result in incorrect results being reported and a delay in treatment.
Code information: Panther Fusion Extraction Reagent-S Lot number: 258402 Panther Fusion Capture Reagent-S Lot number: 257472 Panther Fusion Enhancer Reagent-S Lot number: 257473
Distribution pattern: US: AR, CA, FL, ID, IL, IN, LA, NM, NY, PA, TX, UT, VI, WA, WI OUS: AU, CA
Derived failure modes
-
Unknown
Reagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to invalid runs. This could result in incorrect results being reported and a delay in treatment.