Device product

Vital MIS Spinal Fixation system implants CANN POLY EXT TAB 5.5X50MM, Item Number 810M5550 - Product Usage: are non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Pedicle screw fixation is indicated for skeletally mature patients and for adolescence patients.

Z-2217-2020

May 05, 2020

Class II

Product summary

Firm
Zimmer Biomet, Inc.
Event
Event 85613
Status
Terminated
Classification
Class II
Quantity
3235 total
Official record key
device-enforcement:Z-2217-2020

Official wording

Reason: There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.

Code information: UDI: (01)00887868324451; Lot Numbers: C19H0010 C19I0008 C19J0012 C19K0008 C20A0005 SBM108406 SBM118988 SBM122122