Device product
Omnipod DASH Insulin Management System (mmol/L configuration), Catalog Number PT000011 M/D: INT1D001MM - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
Z-2165-2020
Product summary
- Event
- Event 85552
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1773
- Official record key
device-enforcement:Z-2165-2020
Official wording
Reason: After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.
Code information: UDI: PT000011 10385082000146; Lot Numbers: L000200, L000201, L000203, L000209, L000210
Distribution pattern: The products were distributed to the following foreign countries: Italy, Netherlands, UK.
Derived failure modes
-
Unknown
After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.