Skip to content
Recall Observatory FDA recall evidence

Device product

JOURNEY II UNI RESECTION PREP TRIAL, RIGHT MEDIAL/ LEFT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035831 SIZE 2, CATALOG NUMBER: REF 74035832 SIZE 3, CATALOG NUMBER: REF 74035833 SIZE 4, CATALOG NUMBER: REF 74035834 SIZE 5, CATALOG NUMBER: REF 74035835 SIZE 6, CATALOG NUMBER: REF 74035836 SIZE 7, CATALOG NUMBER: REF 74035837 SIZE 8, CATALOG NUMBER: REF 74035838 SIZE 9, CATALOG NUMBER: REF 74035839 SIZE 10, CATALOG NUMBER: REF 74035840

Z-2133-2020

January 28, 2020

Class II

Product summary

Firm
Smith & Nephew, Inc.
Event
Event 85433
Status
Terminated
Classification
Class II
Quantity
195 units
Official record key
device-enforcement:Z-2133-2020

Official wording

Reason: The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.

Code information: SIZE 1, CATALOG NUMBER: REF 74035831; LOT/BATCH CODES: 19FM01064R, 19HM18082, 19HM21871, 19JM18534, 19JM21086, 19KM07281, 19KM15577 SIZE 2, CATALOG NUMBER: REF 74035832; LOT/BATCH CODES: 19EM21587R, 19FM11181R, 19HM21899, 19HM21902, 19HM21903, 19JM21089 SIZE 3, CATALOG NUMBER: REF 74035833; LOT/BATCH CODES: 19GM21768, 19HM18085, 19HM18085A, 19JM09477, 19JM18537, 19JM18540, 19JM18541, 19JM18542, 19JM18542A SIZE 4, CATALOG NUMBER: REF 74035834; LOT/BATCH CODES: 19FM11183R, 19JM07474, 19JM11137, 19JM21090, 19KM21513 SIZE 5, CATALOG NUMBER: REF 74035835; LOT/BATCH CODES: 19FM01065R, 19HM18086, 19HM18086A, 19HM18088, 19JM18491, 19JM21091, 19KM15580 SIZE 6, CATALOG NUMBER: REF 74035836; LOT/BATCH CODES: 19FM11184R, 19GM21778, 19JM07477, 19JM21093 SIZE 7, CATALOG NUMBER: REF 74035837; LOT/BATCH CODES: 19FM11185A, 19FM11185R, 19HM19006, 19JM21096, 19KM21504, 19KM21506, 19KM21509 SIZE 8, CATALOG NUMBER: REF 74035838; LOT/BATCH CODES: 19FM11191R, 19FM11191S, 19HM18091, 19HM21040, 19JM18543, 19JM21097, 19JM23714, 19KM15583 SIZE 9, CATALOG NUMBER: REF 74035839; LOT/BATCH CODES: 19FM05578R, 19HM18096, 19HM19658, 19KM05613, 19KM15584, 19KM15586 SIZE 10, CATALOG NUMBER: REF 74035840; LOT/BATCH CODES: 19HM18107, 19HM19004, 19JM23666, 19JM23674, 19JM23678, 19JM23678A, 19JM23680, 19KM12708

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.