Skip to content
Recall Observatory FDA recall evidence

Device product

Dover 100% Silicone Foley Catheter with Temperature 16FR 5CC2W, Product Code: 50516T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.

Z-1986-2020

April 30, 2020

Class II

Product summary

Firm
Covidien Llc
Event
Event 85629
Status
Terminated
Classification
Class II
Quantity
3012 units
Official record key
device-enforcement:Z-1986-2020

Official wording

Reason: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Code information: Lot Numbers: 1908700664, 1918351164

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment