Device product
Perceval Single Use Accessory Kit, Size M, REF: ICV1346, Sterile EO, Rx Only, CE, UDI: 8022057015228 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
Z-2985-2020
Product summary
- Event
- Event 86161
- Status
- Terminated
- Classification
- Class II
- Quantity
- 94 kits
- Official record key
device-enforcement:Z-2985-2020
Official wording
Reason: LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
Code information: Kit Lots: 2002060219, 2002060220
Distribution pattern: US Nationwide distribution including in the states of CA, FL, ID, ME, MN, MO, MS, NC, SD, OH, OR, PA, SD, TN, TX, VA, WA, WI and the countries of Canada, Japan, Germany, Netherland, Switzerland, Sweden, UK, Spain, Italy, France, Belgium, Finland, Greece, Slovenia, Malta, Poland, Romania, Croatia, Hungary, Serbia, Australia, Israel, Georgia, Iran, Kuwait, Oman, Quatar, Singapore, Thailand, Taiwan, Russia, South Africa.
Derived failure modes
-
Unknown
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.