Device product
STAINLESS STEEL REPROCESSING TRAY, 2D CAMERA HEAD, IS4000, Model No. 400499-03 - Product Usage: is designed to enable complex surgery using a minimally invasive approach.
Z-2329-2020
Product summary
- Event
- Event 85556
- Status
- Terminated
- Classification
- Class II
- Quantity
- 16 Trays
- Official record key
device-enforcement:Z-2329-2020
Official wording
Reason: Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential breach of sterility and risk of infection. Certain features on the trays with handling and storage may cause these holes.
Code information: Model No. 400499-03; All Batch Numbers; UDI: 00886874115855
Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of Arkansas, Arizona, Alabama, Alaska, California, Connecticut, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Idaho, Kentucky, Kansas, Louisiana, Missouri, Michigan, Maryland, Mississippi, Minnesota, Montana, Maine, Massachusetts, Nevada, New York, North Dakota, Nebraska, North Carolina, New Jersey, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin, Washington and West Virginia. Multiple countries including Canada.
Derived failure modes
-
Sterility assurance
breach of sterility