Skip to content
Recall Observatory FDA recall evidence

Device product

Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com

Z-1552-2020

April 09, 2019

Class II

Product summary

Firm
QUIDEL CARDIOVASCULAR INC
Event
Event 84698
Status
Terminated
Classification
Class II
Quantity
123 kits
Official record key
device-enforcement:Z-1552-2020

Official wording

Reason: Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Code information: Lot Number: C3470AN Part Number 94413

Distribution pattern: US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.