Device product

Infinix-8000V with Catheterization Table CAT-850B or CAT-860B - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Z-2325-2020

March 03, 2020

Class II

Product summary

Firm
Canon Medical System, USA, INC.
Event
Event 85309
Status
Ongoing
Classification
Class II
Quantity
46
Official record key
device-enforcement:Z-2325-2020

Official wording

Reason: Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Code information: Serial Numbers: A5B1642006, A1B17X2004, A5A1612005, A5B1772009, B5B1962002, A3A1522004, A7B16Y2001, A3A1712011, A5B1792010, A5A1532002, A1A1612003, A3A15Z2007, A5A1552003, A5B16Y2007, A3B1862016, A3B1812013, A3A1662009, A9A16X2004, BTA1922001, A3B1842015, AOB1882001, BMB1942001, AOB1892002, A5B1472001, A5B17Z2011, A3A14X2002, A5A1592004, A3A17Z2012, A5C1832012, A5B16Z2008, A1A1582002, A9A1582002, A1A14X2001, A3A1512003, A7C1812002, ABC18X2002, A3A1592005, UTC14Y2008, A3A15X2006, B5B1942001, A3A1682010, A1C18Z2005, A9A1662003, A3A1492001, A3B1822014, A3A1622008

Distribution pattern: US Nationwide distribution including in the states of PA, AR, OH, NC, CA, IA, GA, IL, AL, WV, LA, MO, FL, MS, TX, AZ, MA, NJ, MD, NY, NV, OK, WI, PR.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.