STAINLESS STEEL REPROCESSING TRAY, 8MM ENDOSCOPE, IS4000, Model No. 400498-02 - Product Usage: is designed to enable complex surgery using a minimally invasive approach.

Z-2328-2020

December 19, 2019

Class II

Product summary

Firm
Intuitive Surgical, Inc.
Event
Event 85556
Status
Terminated
Classification
Class II
Quantity
1757 Trays
Official record key
device-enforcement:Z-2328-2020

Official wording

Reason: Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential breach of sterility and risk of infection. Certain features on the trays with handling and storage may cause these holes.

Code information: Model No. 400498-02; All Batch Numbers; UDI: 00886874115848

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of Arkansas, Arizona, Alabama, Alaska, California, Connecticut, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Idaho, Kentucky, Kansas, Louisiana, Missouri, Michigan, Maryland, Mississippi, Minnesota, Montana, Maine, Massachusetts, Nevada, New York, North Dakota, Nebraska, North Carolina, New Jersey, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin, Washington and West Virginia. Multiple countries including Canada.

Derived failure modes

  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    breach of sterility