Device product

Dimension Enzymatic Creatinine, Siemens Material Number 10471520, UDI Number 00842768030512 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.

Z-2284-2020

May 04, 2020

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 85602
Status
Terminated
Classification
Class II
Quantity
8261
Official record key
device-enforcement:Z-2284-2020

Official wording

Reason: The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Code information: Lot #s FC0324, BB0262, GD0218, GC0184, FB0122 ***Updated 8/4/20*** GF1006 ***Added 9/10/20*** FB1050 ***Added 11/2/20*** EA1102

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.