Device product

PAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices. Curved tip Introducer allows for contralateral access to the iliac artery and selected vasculature. Case Label: PAC Tray/5 Catheter Introducer Tray 8F, QTY 10/Bx, Rx Only, STERILE EO

Z-1938-2020

April 08, 2020

Class II

Product summary

Firm
Argon Medical Devices, Inc
Event
Event 85488
Status
Terminated
Classification
Class II
Quantity
80
Official record key
device-enforcement:Z-1938-2020

Official wording

Reason: Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.

Code information: Lot: 11299245

Distribution pattern: US Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.