Device product
Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P18XTSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Z-1994-2020
Product summary
- Firm
- Covidien Llc
- Event
- Event 85629
- Status
- Terminated
- Classification
- Class II
- Quantity
- 0
- Official record key
device-enforcement:Z-1994-2020
Official wording
Reason: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Code information: Lot Numbers: 1914133564
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment