Device product
Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Z-1987-2020
Product summary
- Firm
- Covidien Llc
- Event
- Event 85629
- Status
- Terminated
- Classification
- Class II
- Quantity
- 0
- Official record key
device-enforcement:Z-1987-2020
Official wording
Reason: Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Code information: Lot Numbers: 1914135864
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment