Skip to content
Recall Observatory FDA recall evidence

Device product

VITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.

Z-2771-2020

July 17, 2020

Class II

Product summary

Firm
Ortho Clinical Diagnostics
Event
Event 86038
Status
Terminated
Classification
Class II
Quantity
2349 sales units
Official record key
device-enforcement:Z-2771-2020

Official wording

Reason: The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.

Code information: Lots 0010 - 0013

Distribution pattern: Worldwide distribution - US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0