Device product
therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121
Z-2290-2020
Product summary
- Event
- Event 85565
- Status
- Terminated
- Classification
- Class II
- Quantity
- 530 kits within expiry
- Official record key
device-enforcement:Z-2290-2020
Official wording
Reason: There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.
Code information: All lots
Distribution pattern: US Nationwide.
Derived failure modes
-
Unknown