Device product
da Vinci X Surgical System, IS4200 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are four USM per da Vinci X surgical system.
Z-1897-2020
Product summary
- Event
- Event 83482
- Status
- Terminated
- Classification
- Class II
- Quantity
- 270 USMs on a combination of 100 unique systems (4 USM per system)
- Official record key
device-enforcement:Z-1897-2020
Official wording
Reason: Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.
Code information: UDI: 00886874114216 Device Listing: D295931 Serial Numbers: SL0085, SL0315, SL0317, SL0324, SL0338, SL0350, SL0351, SL0353, SL0354, SL0357, SL0369
Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.
Derived failure modes
-
Unknown
Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.