Device product
Covidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue. Item Number: TA30V3S
Z-0747-2021
Product summary
- Firm
- Covidien Llc
- Event
- Event 86886
- Status
- Completed
- Classification
- Class II
- Quantity
- 3255 units US; 3574 units OUS
- Official record key
device-enforcement:Z-0747-2021
Official wording
Reason: May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis
Code information: Lot Numbers: P0B1142MY, P0C1515MY, P0E0634MY, P0E1353MY, P0C0492MY,P0D1454MYP0E1352MY,P0G0601Y, P0C0493MY UDI: 10884521043145, 20884521043142
Distribution pattern: Nationwide Foreign: Canada, Albania Armenia Australia Austria Belarus Belgium Bosnia And Herzegovina Cyprus Czech Republic Egypt Estonia France Germany Greece Hungary Iceland Israel Italy Japan Jordan Korea Kuwait Latvia Libyan Arab Jamahiriya Lithuania Malaysia Martinique Morocco Myanmar, Netherlands Norway Pakistan Poland Russian Federation Serbia Singapore Slovakia Spain Sweden Switzerland Taiwan Turkey United Arab Emirates United Kingdom Viet Nam
Derived failure modes
-
Unknown
May contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue damage or other delayed secondary complications including infection, peritonitis, and sepsis