Skip to content
Recall Observatory FDA recall evidence

Device product

VITROS XT 7600 Integrated System Software Versions 3.5.1 and below Product Code:6844461 Unique Identifier: 10758750031610

Z-2484-2020

May 20, 2020

Class II

Product summary

Firm
Ortho Clinical Diagnostics
Event
Event 85762
Status
Terminated
Classification
Class II
Quantity
Total= 353 ( U.S.= 167; OUS 186)
Official record key
device-enforcement:Z-2484-2020

Official wording

Reason: Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances

Code information: Affected Serial Number Range: J76000111-J76000525

Distribution pattern: US Nationwide Distribution Foreign: Country AR AU BR CA CL CN CO DO FR ID IN JP KR MX MY NP PE PH SG TH TT VE VN

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances