BD FACS Sample Prep Assistant (SPA) III - Product Usage: Prep Assistant III is intended to prepare human whole blood for flow cytometric analysis on BD FACSCanto II and BD FACSCalibur flow cytometry systems. Pipetting blood, reagents, and lysing solution using the following previously cleared assays for flow cytometric analysis on BD FACSCanto II flow cytometry systems

Z-2358-2020

July 08, 2019

Class II

Product summary

Firm
Becton, Dickinson and Company, BD Biosciences
Event
Event 85129
Status
Terminated
Classification
Class II
Quantity
467 units
Official record key
device-enforcement:Z-2358-2020

Official wording

Reason: The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.

Code information: Catalog (Ref) No. 647205

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AL, CO,CT, GA, IL, MA, MI, MD, NC, NH,NJ, NY, OH, PA, TN, TX, VA, and VT. The countries of Australia, Botswana, Brazil, Canada, Chile, China, Colombia, Costa Rica, Europe, Hong Kong, India, Italy, Jamaica, Japan, Kazakhstan, Malaysia, Martinique, Mexico, New Zealand, Nicaragua, Peru, Puerto Rico (US), Romania, Russia, Singapore, South Africa, South Korea, Swaziland, Switzerland, Turkey, US, Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The recalling firm confirmed that SPA II and III users may encounter probes with unacceptable wear prior to reaching the 2500 piercings guideline indicated in the Instructions for Use (IFU). When the probe is worn, particles from the rubber cap are more likely to clog the probe, resulting in blocked or inadequate sample dispense.