Skip to content
Recall Observatory FDA recall evidence

Device product

Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Z-2164-2020

April 02, 2020

Class II

Product summary

Firm
Insulet Corporation
Event
Event 85552
Status
Terminated
Classification
Class II
Quantity
400
Official record key
device-enforcement:Z-2164-2020

Official wording

Reason: After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.

Code information: UDI: PT - 000010 - 10385082000139; Lot Numbers: L000202

Distribution pattern: The products were distributed to the following foreign countries: Italy, Netherlands, UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.