Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001

Z-0473-2021

October 26, 2020

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 86718
Status
Terminated
Classification
Class II
Quantity
US 392 units; OUS 449 units
Official record key
device-enforcement:Z-0473-2021

Official wording

Reason: Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values

Code information: Software version: V1.23.1 or lower UDI: 00630414007960

Distribution pattern: Worldwide distribution - US Nationwide distribution and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Peru, Poland, Portugal, Qatar Republic, Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay, Vatikancity, Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values