Skip to content
Recall Observatory FDA recall evidence

Device product

Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The device is a component of the MPS Console used as replacements for worn or damaged MPS Console top covers)

Z-2107-2020

December 17, 2019

Class II

Product summary

Firm
Quest Medical, Inc.
Event
Event 85389
Status
Terminated
Classification
Class II
Quantity
5 units
Official record key
device-enforcement:Z-2107-2020

Official wording

Reason: Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention.

Code information: Lot Number: 059135

Distribution pattern: US Nationwide distribution including in the states of NY, PA, TN, WV. OUS: None

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention.