Skip to content
Recall Observatory FDA recall evidence

Device product

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.

Z-2136-2020

January 29, 2020

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 85467
Status
Terminated
Classification
Class II
Quantity
2738 vials
Official record key
device-enforcement:Z-2136-2020

Official wording

Reason: Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.

Code information: Lot Number: M909206

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries of CANADA, JAPAN, LEBANON, SAUDI ARABIA, TAIWAN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.