Device product
REF 27219-1 X1, Medtronic, Delta Valve, Neonatal, Performance Level 1.0, CE 0344, Sterile EO - Product Usage: Neonatal Delta Snap Shunt Assemblies are components of CSF-Flow Control Shunt Systems designed to provide controlled CSF flow from the ventricles of the brain into the peritoneal cavity.
Z-0871-2021
Product summary
- Event
- Event 87069
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 95 units
- Official record key
device-enforcement:Z-0871-2021
Official wording
Reason: Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction
Code information: Lot # 0219927059, 0220149614, 0220393237, 0220601648,0220601649, 0220813029
Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IL, KS, KY, LA, MA, MI, MO, NV, NM, NY, ND, OH, OK, TN, VA and the countries of Albania, Australia, Brazil, Cyprus, Ireland, Mexico, Qatar, Romania, Russian Federation, Serbia, South Africa, Spain, United Kingdom.
Derived failure modes
-
Unknown
Neonatal Valves were manufactured with a ventricular catheter connection site shorter than the design specification. The defect can lead to difficulty in securing the valve to the proximal catheter which could result in disconnection of the ventricular catheter from the valve and shunt malfunction