Skip to content
Recall Observatory FDA recall evidence

Device product

Critical Care Decontamination System (CCDS) Compatible N95 respirators - Product Usage: Used in decontaminating compatible N95 respirators2 for multiple-user3 reuse by healthcare personnel (HCP)4 to prevent exposure to pathogenic biological airborne particulates when there are insufficient supplies of Filtering Facepiece Respirators (FFRs) resulting from the Coronavirus Disease 2019 (COVID-19) pandemic.

Z-0309-2021

August 13, 2020

Class II

Product summary

Firm
Battelle Memorial Institute
Event
Event 86347
Status
Terminated
Classification
Class II
Quantity
898
Official record key
device-enforcement:Z-0309-2021

Official wording

Reason: Masks processed at one site were not maintained at levels of condensation during a portion of the decontamination cycle.

Code information: All decontaminated compatible N95 respirators distributed from Providence, RI between 07/01/2020 and 08/12/2020.

Distribution pattern: US Nationwide distribution including in the states of CT, MA, RI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Masks processed at one site were not maintained at levels of condensation during a portion of the decontamination cycle.