Device product
IntelliVue MX700 patient monitor, model no. 865241
Z-2697-2020
Product summary
- Event
- Event 85908
- Status
- Terminated
- Classification
- Class II
- Quantity
- 28
- Official record key
device-enforcement:Z-2697-2020
Official wording
Reason: Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Code information: Serial numbers DE58562415 DE58562344 DE58562353
Distribution pattern: Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
Derived failure modes
-
Unknown
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.