Device product

RayStation standalone software treatment planning system (RayStation 8A, 8B, 9A, and 9B, including all service packs) model no. 8.0, 8.1, 9.0, 9.1, 9.2

Z-2426-2020

May 26, 2020

Class II

Product summary

Firm
RAYSEARCH LABORATORIES AB
Event
Event 85796
Status
Ongoing
Classification
Class II
Quantity
12
Official record key
device-enforcement:Z-2426-2020

Official wording

Reason: It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.

Code information: Software versions 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 and 9.2.0.483

Distribution pattern: Distributed to one consignee in Nashville. Foreign distribution to one consignee in Belgium.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.