MEDICREA LIGAPASS TENSION PULLEY HANDLE XS- intended to provide thoracic and/or lumbar posterior stabilization for degenerative pathologies Ref : A08200200

Z-0463-2021

October 08, 2020

Class II

Product summary

Firm
Medicrea International
Event
Event 86657
Status
Terminated
Classification
Class II
Quantity
17 units US; 8 units OUS Expanded: 1 US and 3 OUS
Official record key
device-enforcement:Z-0463-2021

Official wording

Reason: Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,

Code information: Lot Numbers: US: 17E0550-1, 17E1089R/2, 18E0278, OUS: 19B0962, 19B0595, 19B0601 Expanded Recall: 11/3/20 17E0550/2, 19B0602, 19B0963 and 19B0964

Distribution pattern: CA Foreign:France, Belgium

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0