Cobra x5 1-CH EMG ET Tube, 6.0mm, REF: LTE700S-5, Rx Only, Sterile EO UDI: B006:TE700S52$$7050918B - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Z-2903-2020

July 21, 2020

Class II

Product summary

Firm
Neurovision Medical Products Inc
Event
Event 86102
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-2903-2020

Official wording

Reason: The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Code information: Kit Lots: 073119A 082019B 022820A 041620C 062520A

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.