Skip to content
Recall Observatory FDA recall evidence

Device product

ALTERA Spacer ALTERA Spacer, 10x36, 10-14mm, 8¿

Z-2867-2020

July 22, 2020

Class II

Product summary

Firm
Globus Medical, Inc.
Event
Event 86164
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-2867-2020

Official wording

Reason: Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.

Code information: Part #: 1124.1213; Lot # / UDI #: BAX128PE / (01)00889095069891(10)BAX128PE; BAX207CE / (01)00889095069891(10)BAX207CE; BAX128NE / (01)00889095069891(10)BAX128NE; BAX250EE / (01)00889095069891(10)BAX250EE; BAX256BE / (01)00889095069891(10)BAX256BE

Distribution pattern: Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.