Device product
Bioseal Trachea Extender 1/pl 50pk/Cs, REF TRAX01/50, LOT XXXX, STERILE EO, Single Use Only, Date of Manufacture XXXX, Expiration Date XXXX, LF, UDI: (01) 00630094916019 - Product Usage: Airway connectors intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or mask.
Z-2795-2020
Product summary
- Event
- Event 85288
- Status
- Terminated
- Classification
- Class II
- Quantity
- 900 units
- Official record key
device-enforcement:Z-2795-2020
Official wording
Reason: The reported stress crack in the port of the swivel elbow where the flip cap is inserted could result in leakage from the breathing system resulting in the prescribed ventilation not being delivered to the patient. These cracks could possibly leak if they reach a significant size and pose a serious heath risk to patient.
Code information: Bioseal Part # TRAX01/50 Bioseal Lot #'s 400453, 400456, 400461, 400472, 400476
Distribution pattern: US Nationwide distribution including in the state of CA.
Derived failure modes
-
Unknown
The reported stress crack in the port of the swivel elbow where the flip cap is inserted could result in leakage from the breathing system resulting in the prescribed ventilation not being delivered to the patient. These cracks could possibly leak if they reach a significant size and pose a serious heath risk to patient.