Match Point System The Match Point System is intended to by uses as a surgical instrument to assist in the intraoperative positioning of glenoid components used with total and reverse shoulder arthroplasty by referencing anatomic landmarks of the shoulder that are identifiable on preoperative CT-image scans.

Z-2886-2020

July 17, 2020

Class II

Product summary

Firm
Materialise N.V.
Event
Event 86113
Status
Terminated
Classification
Class II
Quantity
128
Official record key
device-enforcement:Z-2886-2020

Official wording

Reason: The procedure side indicated in the top header of the Shoulder Case Planning Report incorrectly indicated the procedure side as "Left" for surgical plans for a "right" procedure side.

Code information: ID #s: 804-03-057, 804-25-148, 804-03-058, 804,07,022 UDI #s: E314804030570, E314804251480, E314804070220

Distribution pattern: Nationwide distribution to: US including PR and International to UK and Australia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The procedure side indicated in the top header of the Shoulder Case Planning Report incorrectly indicated the procedure side as "Left" for surgical plans for a "right" procedure side.