Device product
NA-201SX-4022-C ASPIRATION NEEDLE ViziShot 22G w/o stylet, model no. NA-201SX-4022-C - Product Usage: Used with ultrasound endoscopes for ultrasound guided fine needle aspiration of submucosal and extramural lesions of the tracheobronchial tree and esophagus.
Z-1062-2021
Product summary
- Event
- Event 87074
- Status
- Terminated
- Classification
- Class II
- Quantity
- 7,400,000 (globally); 260,395 (US)
- Official record key
device-enforcement:Z-1062-2021
Official wording
Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Code information: 03K, 04K, 95V, 96V, 97V, 98V, 99V, 9XV, 9YV, 9ZV, 01V, 02V, 03V, 04V, 05V, 06V, 07V, 08V, 09V, 0XV
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Labeling or packaging
inadequate package
-
Sterility assurance
sterile packages with inadequate