Device product
139" (353 cm) 15 Drop Admin Set w/3 MicroClave Clear, 4-Way Stopcock, 3-Port NanoClave Manifold w/Check Valve, Spin Luer, 2 Ext. 1 unit per pouch, 25 pouches per case. UDI:(01)10887709121147(17)250101(30)25(10)4556713; (01)10887709121147(17)250101(30)25(10)4603383; (01)10887709121147(17)250101(30)25(10)4573356. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).
Z-0233-2021
Product summary
- Event
- Event 86054
- Status
- Terminated
- Classification
- Class II
- Quantity
- Total of all products (Listed #1 thru 101) = 304735 units
- Official record key
device-enforcement:Z-0233-2021
Official wording
Reason: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Code information: Lot Numbers:4556713, 4603383, 4573356.
Distribution pattern: Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Derived failure modes
-
Manufacturing or process control
manufacturing defect