Skip to content
Recall Observatory FDA recall evidence

Device product

SD-400U-10 SNAREMASTER PLUS Hot/Cold 10m, model no. SD-400U-10 - Product Usage: Used with endoscopes for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.

Z-0991-2021

January 04, 2021

Class II

Product summary

Firm
Aomori Olympus Co., Ltd.
Event
Event 87074
Status
Terminated
Classification
Class II
Quantity
7,400,000 (globally); 260,395 (US)
Official record key
device-enforcement:Z-0991-2021

Official wording

Reason: Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

Code information: 97V, 98V, 9XV, 9YV, 03V, 05V, 06V

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    inadequate package
  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    sterile packages with inadequate