Device product
PROXIMALE, 28 cm (11 IN) APPX 1.0ml, EXT Set, 6-PORT NanoClave MANIFOLD, NanoClave. 1 unit per pouch, 50 pouches per case. UDI:(01)10840619093195(17)250101(30)50(10)4600271; (01)10840619093195(17)241201(30)50(10)4537220. Administration sets and their accessories are single use, pre-sterilized devices that provide access for the administration of fluids from a container to a patients vascular system through the administrations set needle or catheter, which is inserted into the vein
Z-0273-2021
Product summary
- Event
- Event 86054
- Status
- Terminated
- Classification
- Class II
- Quantity
- Total of all products (Listed #1 thru 101) = 304735 units
- Official record key
device-enforcement:Z-0273-2021
Official wording
Reason: Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Code information: Lot Numbers:4600271, 4537220.
Distribution pattern: Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Derived failure modes
-
Manufacturing or process control
manufacturing defect